Model No.︰LDN-Q16
Brand Name︰CSOIF
Country of Origin︰China
Unit Price︰US $ 5800 / pc
Minimum Order︰1 pc
Product introduction:
1. Used for the detection of organophosphorus and carbamate pesticide residues in fruits, vegetables and other agricultural products.
The enzyme inhibition rate method in national standards is a sensitive, efficient and accurate method for the rapid detection of pesticide residues in China.
2. Automatic pesticide residue detector can realize the full automation of rapid detection of pesticide residue, with built-in detection reagent such as enzyme, substrate and color reagent, etc. During detection, only samples to be tested need to be put in, and reagent addition, mixing and temperature bath can be completed automatically, which can realize accurate and rapid detection of pesticide residue.
In the detection process, the temperature is constant, the reagent is quantitative, and the automatic cleaning can be realized after the detection process and use, which is simple, fast, stable and reliable, eliminating the error introduced by human factors in the traditional detection.
Ii. Summary of advantages
1. Convenient operation: integrated design of the whole machine, integrated main control system, human-computer interaction system, micro thermal printer, single fully functional independent operation, no need for other peripherals.
2. Technical advantages: the unique patented sample adding and mixing technology is adopted to add samples independently with various reagents. There is no contact between sample adding needles and samples during the whole process, so the possibility of sample cross contamination is completely eliminated.
3. Environmental protection: there is no need to clean pipelines and sample adding needles, no need to produce cleaning liquid and no waste liquid, and the detection process is more environmentally friendly.
4. Full closed-loop control: a variety of sensors are built in to automatically identify whether samples exist or not, high-precision closed-loop temperature control, accurate drip addition of reaction reagent, and timely feedback of reagent stock.
5. High detection accuracy: single wavelength interference filter is adopted to eliminate the influence of spectral drift of the light source;
Synchronous detection of reference channel signals, automatic correction of light source fluctuations, and the use of independent research and development of signal detection algorithm, greatly improve the accuracy of the instrument.
Iii. Technical parameters and advantages
1. Equipped with 16 detection channels and high-performance control algorithm, the instrument can automatically identify and locate detection channels, which can be used simultaneously or separately;
2. The detection channel is equipped with automatic cleaning system, with less water consumption;
3. Fully automatic addition of standardized pesticide residue reagents without manual intervention;
4. The special patented sample adding technology is adopted. Three specially customized independent reagent needles are used to ensure that the reagent needles do not touch the sample solution.
5. Reagent needle: fixed installation, no up and down, reciprocating movement, good stability and reliability;
6. Adopt high speed shock mixing technology, make sample liquid automatically mix, the mixing degree can reach more than 97%, ensure the full and complete process reaction;
7. High precision pipet pump and high sensitivity signal sensor are adopted to monitor the reagent addition process and ensure the volume accuracy of reagent addition;
8. High sensitivity signal sensor is adopted to monitor the residual reagent quantity and ensure the sample adding reliability;
9. The high precision temperature detection circuit, coupled with reliable control algorithm, makes the temperature stable in the range of 37±0.5℃, ensuring the accuracy and stability of the detection results.
10. Detection light source: LED cold light source, no need to replace the light source throughout its life. Advanced electrical structure and optimized control algorithm are adopted to ensure the stability of the output power of the detection light source;
★ 7-inch capacitive touch screen, using Android intelligent operating system, simple and easy to use;
12. Information function, with wi-fi communication function;
13. Integrated thermal printer, instant output of inspection report;
14. Detection speed: 50T/H;
15. Repeatability: CV≤3%
16. Absorbance test range: 0 ~ 3A;
17. Absorbance test accuracy: ±1% (0 ~ 2A);
18. Absorbance test repeatability: ±0.01A (0 ~ 2A);
19. Stability of absorbance test: ≤0.01A;
20. Power supply: 100-240vac, 50/60hz, 40W;
21. Weight: <8Kg;
22. Size: 430mm×250mm×230mm
Payment Terms︰ TT